Input specification
|
Input connector |
BNC(C02 type), terminal block |
Input format |
Single-phase, AC/DC/non-voltage selectable(+12 V pull-up for open collector devices)
Two-phase signal with 90° phase difference(DC input only) |
Input voltage |
AC input signal voltage range :0.3 Vp-p to 30 Vp-p |
DC input signal voltage range :Lo;+1 V or less, Hi;+4 to 30 V |
Filter |
OFF / 20 kHz / 120 kHz low-pass filter |
Input impedance |
2.2 kΩ |
AC coupling frequency |
0.9 Hz approx. -3 dB |
Minimum input pulse width |
1.5 µs |
Input signal detection edge |
Rising / falling selectable |
Input frequency range |
0.2 Hz to 320 kHz
For full scale mode output |
Can be set from 1 to 320000 Hz in increments of 1 Hz |
Can be set from 1 to 320000 r/min in increments of 1 r/min |
Can be set from 1 to 320000 m/min in increments of 1 m/min |
For deviation mode |
Selectable from the measurement frequency in the range up to 320 kHz
±1 %, ±5 %, ±10 %, ±20 %, ±50 %, ±100 % or ±1 to 180,000 (can be set in increment of 1Hz, 1r/min, or 1m/min.) |
|
Signal output specification
|
Analog output signal
|
Analog output connector |
BNC/C02 type(voltage output)or terminal block (phoenix contact:MC1,5/6-STF-3,81)(voltage output)selectable |
Conversion time |
1 cycle +3.5 µs or less |
D/A resolution |
16-bit |
Filter |
OFF / 3 Hz / 10 Hz / 1 kHz selectable, low-pass filer |
Analog output connector signal |
Voltage output |
0 to 10 V(Full scale mode output, direction recognition function OFF)
± 5 V(Full scale output mode, direction recognition function ON, deviation output mode, automatic center frequency follow-up mode (option :FV-0151)) |
± 10 V(Deviation output mode: option FV-0153) |
Load resistance 100 kΩ or more |
Current output |
0 to 16 mA / 4 to 20 mA |
0 to 16 mA, load resistance 500 Ω or less (at the time of shipment) |
|
Output DC offset |
Voltage output:± 10 mV or less |
Current output:± 0.01 mA or less |
Temperature coefficient |
Voltage output:± 0.02 %/F.S./℃ |
Current output:± 0.07 %/F.S./℃ |
Output impedance |
50 Ω or more(voltage output) |
Load resisntance |
Voltage output:100 kΩ or more |
Current output:500 Ω or less |
Linearity (For full scale mode) |
Voltage output |
DC:± 0.1 %(to 180 kHz) ± 0.2 %(to320 kHz) |
AC:± 0.2 %(to 180 kHz) ± 0.4 %(to320 kHz) |
Current output |
DC:± 0.7 %(to 180 kHz) ± 1.4 %(to320 kHz) |
AC:± 1.4 %(to 180 kHz) ± 2.8 %(to320 kHz) |
|
Pulse output signal
|
Pulse output |
BNC(C02 type)
Output connector voltage: Lo: +1 V or less, Hi:+4.5 V or more
Load resistance:100 kΩ or more |
Pulse output (option: FV-0154) |
Open collector output
Withstand voltage:DC30 V, sink current:50mA or less |
Comparator output signal(option: FV-0152)
|
Comparator output terminal |
Terminal block (phoenix contact MC1,5/6-STF-3,81) |
Output format |
Semiconductor relay (one make contact) 2ch |
Maximum contact capacity |
DC 30 V/1 A |
Contact ON resistance |
25 Ω or less |
Setup |
UPPER, LOWER
Relay ON: UPPER setup value ≤ measurement value
Relay ON: LOWER setup value > measurement value |
Other functions
(provided as standard) |
Delay function |
The delay function performs comparator output when the set value is consecutively exceeded the preset number of times. Can be set in the range from OFF (0) to 100 times. |
Hysteresis function |
Return values of frequency, rotation seed or moving speed can be set in the range from 0 % to 20 % in increments of 1 %. |
Hold function |
This function sets the minimum output time to keep contact ON status in the range from 1 to 1000 ms, in increments of 1 ms. |
Function
|
Moving average |
OFF / simple moving average 2 to 720 times / exponential moving average 2 to 64 times |
Number of pulses |
1 to 999999 P/R |
Factor |
0.0001 to 99999 |
Multiplying / dividing |
Multiplying:OFF / 2 / 4 (two-phase signal with 90° phase difference) frequency dividing: OFF, 2 to 1000 |
Condition memory |
Can be stored up to 10 types of conditions. |
Rapid deceleration follow-up function(Estimating calculation) |
OFF / TYPE1 / TYPE2 |
Stop estimating calculation time |
OFF / NUM:1.5 times to 16 times / TIME:1 ms to 4000 ms |
Direction recognition |
OFF / ON (two-phase signal with 90° phase difference) |
CAL |
Voltage output ZERO : 0 V, FULL: 10 V |
Current output ZERO : 0 mA (4 mA), FULL:16 mA (20 mA) |
Display
|
Display accuracy |
displayed value* x (±0.01%) ± 1 count or less
(* Count value : The value that a decimal point is ignored. ) |
Display |
Fluorescent display tube (display range 69.85 mm × 11.45 mm)
Selectable from four brightness degrees |
Display update time |
0.5s, 1s |
Display unit |
Hz, r/min, m/min, USER |
Status LED display |
SIG |
Flashes each time a signal is input |
Measurement mode |
Full scale, deviation, automatic center frequency follow-up function(option: FV-0151) |
Sensor power
|
Sensor power |
+12 V±1 % 150 mA / +5 V±10 % 150 mA
Selectable by SENSOR/POWER switch on the real panel. |
Conforming standard
|
CE marking |
Low Voltage Directive 2006/95/EC EN61010-1:2010 |
EMC Directive 2004/108/EC EN61326-1:2006 Class A Table 2 |
General specification
|
I/O ground |
Signal ground of V OUT and I OUT are in common.
Signal ground of SIG IN, SIG 1, SIG 2 and sensor power (+12V, +5V) are in common.
The signal input and sensor power ground shall be isolated from the chassis.
The input and output grounds shall be isolated from each other. The withstand voltage between the isolated signal grounds and between chassis shall be up to 42 Vpk. |
Power requirement |
16 VCD ±10 %
Exclusive AC adapter (100 to 240 VAC) |
Power consumption |
Approx. 36 VA or less |
Operating temperature range |
0 to 40 ℃ |
Storage temperature range |
-10 to 50 ℃ |
Operating humidity range |
5 to 80 %(with no condensation) |
Storage humidity range |
5 to 85 %(with no condensation) |
Outer dimensions |
210(W)×44(D)×200(D) mm(excluding protruded section) |
weight |
Approx. 1 kg |
Accessory |
Exclusive AC adaptor (PS-P20023B / VM1391-VM1700 2m) X 1, instruction manual X 1, connector (MC1.5/6-STF-3.81) X 1 |
Option
|
Option |
FV-0151: Automatic center frequency follow-up function
FV-0152: Comparator output function
FV-0153: Deviation scale change function(±10 V)
FV-0154: Open collector output function(pulse output)
FV-0014: Panel mounting fixture |