| Model | YS6002 | 
			
				| Optical Geometry
 | Transmittance:d/0 (diffused illumination, 0-degree viewing angle); similar to CIE15 standard 0/0(parallel illumination, 0-degree viewing angle);
 Confirms to the standard of ASTM D1003/1044, CIE No.15, GB/T 3978, GB
 2893, GB/T 18833, ASTM E308, DIN5033 Teil7
 | 
			
				| Application
 | wildly used in the field of glass processing, plastic processing, film manufacturing, display screen processing and packaging industry, liquid
 pharmaceutical analysis, laboratory etc.
 | 
			
				| Intergrating Sphere Size | Φ154mm | 
			
				| Light Source | 360~780nm Combined LED Lamp | 
			
				| Spectral Separation Device
 | Concave-Grating | 
			
				| Detector | 256 Image Element Double Array CMOS Image Sensor | 
			
				| Wavelength Range | 360~780nm | 
			
				| Wavelength Pitch | 10nm | 
			
				| Half Bandwidth | 10nm | 
			
				| Reflectance Range | 0~200% | 
			
				| Measurement Aperture | Transmittance: Φ25.4mm | 
			
				| Color Spaces | CIE LAB,XYZ,Yxy,LCh,CIE LUV,Musell,s-RGB,HunterLab,βxy,DIN Lab99 | 
			
				| Color Difference Formulas
 | ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE(Hunter) | 
			
				| Other Colorimetric Data
 | Haze (ASTM D1003/1044),transmittance rate T, WI(ASTM E313,CIE/ISO,AATCC,Hunter),
 YI(ASTM D1925,ASTM 313),
 TI(ASTM E313,CIE/ISO),
 Mt(Matamerism Index ),
 Color Strength, Staining fastness,Color Strength, Opacity,
 Pt-Co index and Gardner index,Color code (ASTM D1500)
 | 
			
				| Observer Angle | 2°/10° | 
			
				| Illuminant | D65,A,C,D50,D55,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,DLF,TL83,TL84,TPL5,U30 | 
			
				| Display Data | Spectral/Values, Chromaticity Values, Color Difference Values/Graph,Color Diagram, Color Simulation, Pass/ Fall Result, Color Offset
 | 
			
				| Measurement Time | About 1.5s | 
			
				| Repeatability
 | Spectral transmittance: for test transmission reference white within ΔE*ab 0.05% (400~700nm: within 0.04%);
 Chromaticity Data: Φ25.4mm,within ΔE*ab 0.03 (after do warm-up and
 calibration, it is the average value to measure transmission reference white
 plate for 30 times at 5 second intervals)
 | 
			
				| Inter-instrument Agreement
 | Φ25.4mm,within ΔE*ab 0.2 (average value for 5 pcs standard transmittance color plates measurement )
 | 
			
				| Size | 370X300X200mm(L*W*H) | 
			
				| Weight | About 9.6kg | 
			
				| Power | DC 24V,3A Power Adapter | 
			
				| Illuminant Life Span | 5 years, more than 3,000,000 times measurement | 
			
				| Screen | 7-inch TFT Capacity Touch Screen Display | 
			
				| Interface | USB,Printing Port | 
			
				| Storage | Standard:5000; Sample:20000 | 
			
				| Language | Simplified Chinese, Traditional Chinese, English | 
			
				| Working Temperature | 0~40℃(32~104°F) | 
			
				| Storage Temperature | -20~50℃(-4~122°F) | 
			
				| Standard Accessory
 | Power Adapter, User Manual, QC software(please download from website), USB cable, reflectance white and black board, transmittance black board,
 25.4mm aperture, transmissive test component
 | 
			
				| Optional Accessory | Miniature- printer | 
			
				| Note | The specifications are subject to change without notice. |