Infrasonic Meter TYPE 7144/3348D is ultralow-frequency sound pressure level meter to measure infrasonic or aerial vibration generated by explosive eruption of volcanoes.Measurement of ultralow-frequency sound pressure is possible in the range 2Pa~632Pa and 0.1~100Hz.

Reduction of input impedance has been realized by the employment of ceramic sensor, which has then offered an environmental suitability such as long-term outdoor measurement.

Case Studies

Aerial vibration is rapid pulsation of air generated around the crater by the explosive eruption of volcanoes, which thus travels as low-frequency sound pressure.
Due to ultralow-frequency sound pressure, the aerial vibration can not be heard by human ear but easily vibrates the building walls or windows during the propagation, sometimes giving a big damage such as breakdown of windows.
Infrasonic Meter means air vibrations are generated when a rapid change in barometrical pressure happens in the crater because of the eruption of the explosion of the volcano, and the impact spreads as a sound wave.

Volcano SHINMOEDAKE viewed from our Miyazaki factory.

TYPE 7144

Installation site of Infrasonic Meter

7144_3348d_secchi

Product Specifications

【Specification】Infrasonic Meter TYPE 7144

Frequency range 0.1~100Hz
Sensitivity(including Preamp) 3.5mV/Pa
Dynamic range 0.0632Pa~632Pa(Effective value) (When each range is full-scale)
Sensor Ceramic type
Preamplifier Charge converter
Size Φ23.77mm × 130mm

【Specification】Amplifier TYPE 3348D

Input Channel 1Channel
Terminal 6P connector(R03-R6F made by TAJIMI)
A:+15V B:-15V C:SIG E,F:GND
Output Terminal BNC connector (Digital:USB command)
Voltage Real-time waveform output ±1V/F.S.(632Pa)
Digital 16Bit (±32767/F.S.)
Noise Level 2mVrms or less
Measurement Range 2Pa~632Pa(Effective value)
Range 2Pa/6.32Pa/20Pa/63.2Pa/200Pa/632Pa 6stages
Gain stability level Within ±1%(Output terminal) Full-scale
Linearity Within ±1%(Output terminal) Full-scale
Phase lag ±2°(Positive aspect)
Indicator LCD:16 characters,2lines Display cycle:1 second
Display of Measurement value, Low-pass filter, Level range
Distortion of higher harmonics wave 1%
Frequency response 0.1Hz~100Hz ±3dB (when Low-pass filter OFF)
Dynamic response time constant 1 second (Slow),10 seconds
Low-pass filter 100Hz -3dB Attenuation feature -12dB/oct
Electric calibration Electric calibration by built-in oscillator(sine wave of 80Hz)
I/O terminal Control and digital output (USB1.1) of Infrasonic Meter by PC
Power DC+12V ±10% Terminal stand M3 or AC adapter
Power consumption Approx.1.5W(Approx.115mA) or less
Operating temperature limit -20℃~50℃ / 90% or less(Dewfall strict prohibition)
Size and Weight 160(W)×94(H)×230(D)